The MI-TS-3871 from Gustav Klein is a multi channel infeed test system designed for the testing and emulation of batteries, fuel cells, and powertrains.
It provides a high performance output of up to 1200V and 1200A, making it suitable for demanding applications. The system delivers a maximum power output of 650kW in single operation, which can be increased to 1.3MW when operated in parallel, offering the flexibility required for larger and more complex testing setups.
Up to 756kVA+
350Vac L-N/606Vac
L-L or ±500Vdc
Up to 1,200Hz
| Specification | Details |
|---|---|
| Input voltage / input frequency |
380 / 400* / 440 / 480 / 500 / 690 V ±10% 3-phase, PE 50 / 60 Hz ±5% |
| Measurement resolution |
Voltage: 16-bit ADC Current: 16-bit ADC |
| Control accuracy |
Voltage: 0.1% fs Current: 0.1% fs |
| Voltage ripple | ≤ 0.1% effective fs |
| Current ripple | ≤ 0.1% effective fs |
| Dynamic voltage tolerance (0–100% nominal in 3 ms) |
Battery tester < 3% Battery simulator < 1% |
| Short-circuit behavior |
Systems up to 600A: Icw < 3 kA Systems > 600A: Icw < 8 kA |
| Minimum distance from ceiling | 300 mm (Standard, IP20) |
| Permissible environmental conditions |
Storage: EN 60721-3-1 Transport: EN 60721-3-2 Operation: EN 60721-3-3 1K21 / 1M11, +5 to +40 °C 2K12 / 2M4, -45 to +70 °C 3K22 / 3M11, +5 to +40 °C |
| Lineup |
Workplace with restricted access Installation on non-combustible flooring |
| Protection class |
IP20 according to IEC 60529 (others available on request) |
| Security | EN ISO 13849-1 |
| Basic standard | EN 62040 |
| EMC |
EN 61000-2-4 (Mains disturbances) EN 61000-6-2 (Immunity) EN 61000-6-4 (Emission) EN 61800-3 Cat C2 (A1) (Variable speed electric drives) |
CALTEST INSTRUMENTS PVT LTD